Free Download Distek Evolution 4300 Manual Programs For Single
REASON: The circulator is not turning on due to thermal switch reset. This occurs when the TCS pump is not primed properly. PRE-REQUISITE: Service Engineer or person performing this service must have proper training in servicing the instrument. RESETTING THE THERMAL SAFETY SWITCH Step 1. Turn off the TCS and unplug from the AC outlet prior to resetting the thermal switch. Make sure that the TCS has had time to cool down prior to resetting the thermal safety switch. This can take up to 10 to 20 minutes.
Includes manual, (2). Home » Products » Chemistry » Dissolution System » Distek Evolution 4300 Dissolution System. Used Distek Evolution 4300 Dissolution. Filed under Cipher Software, Evolution 4300, Evolution 6100, Evolution 6300. The Evolution 6100 is Distek’s flagship. Create a free website or blog. Free Download Distek Evolution 4300 Manual Programs. Download user guide, user manual. E-Learning programs for the Lab. Evolution 4300 autosampler and Opt.
On the back of the TCS there is a small access hole where the thermal safety switch is located (As shown below). Using a non-metallic tool or the end of a pen or pencil, reset the thermal safety switch by depressing the button. You should feel or hear a click indicating that the switch has been reset. Once the thermal safety switch has been reset, make sure the TCS has been properly primed or the thermal safety switch will trip again. Please see Priming the TCS Pump steps below. Turn on the unit and enable the heater.
Information about resetting the thermal safety switch can be found on page 3-14 in the 2100C Operation Manual. Information about priming the pump can be found on page 2-6 in the 2100C Operation Manual and on page 2-8 of the EVO 6300 Operation Manual. Custom And Vintage Sdx Keygen For Mac. PRIMING THE TCS PUMP Step 1. Attach a 24 inches (61 cm) long x ½ inch (13mm) ID drain tubing to drain valve. Raise the drain tubing to vertical position as shown.
Drain valve should be tilted up. Cover the end of the tubing with your thumb and open the drain valve.
Slowly lower the tubing while slowly opening the end of the tubing allowing water to fill up as much of the tubing. Once filled, close again the end of the tubing with your thumb and raise the tubing. Uncover the end of tubing to force water in to the pump inlet. Perform this step at least two to three times. Turn on the TCS circulator and observe that air has been purged from the pump indicating that the pump was primed correctly. If it does not, repeat step 2 and 3 until all the air is purged from the pump. When connecting a Distek dissolution system to the Evolution 4300 syringe-pump dissolution autosampler, you will need to connect an RS-485 cable between the two systems to allow for electronic control of the dissolution unit from the autosampler.